Information about the Zeiss Cell Discoverer 7. FeaturesWidefield, confocal and Airyscan imaging with combined modes possibleAccepts almost all sample types from slides to well platesFull temperature control from 5 to 45°C with rapid variationCO2 up to 11% and nitrogen hypoxia moduleEasy magnification switching from 2.5x up to 100x magnificationGuided Acquisition module enabling low resolution overview, automated analysis and detailed high-resolution acquisition of detected features.Accepts AI models and integrates with Arivis cloud and Arivis Pro for model training.ObjectivesEach can be adapted with 0.5, 1 and 2x magnificationMagnificationNA (0.5x/1x/2x)ImmersionWorking DistanceX50.12/0.25/0.35Air5.1mmX200.35/0.7/0.7Air2.2mmX200.5/0.8/0.95Air0.76mmX501.2/1.2/1.2Water0.84mmWidefield Filter SetsExample FluorophoreExcitation LEDDichroicEmission FilterDAPI, Alexa 405, Hoechst 33258385QBS 405+493+575+653BP 425/30QBP 425/30 + 514/30 + 592/25 + 709/100eCFP, Lucifer Yellow, Alexa 430423TBS 450+538+610TBP 467/24 + 555/25 + 687/145eGFP, Alexa 488, FITC469QBS 405+493+575+653514/30QBP 425/30 + 514/30 + 592/25 + 709/100eYFP, TOTO-1508TBS 450+538+610TBP 467/24 + 555/25 + 687/145Cy3, TRITC, DsRed, tdTomato555QBS 405+493+575+653592/25QBP 425/30 + 514/30 + 592/25 + 709/100mCherry, Alexa 568555(590)TBS 450+538+610681/45TBP 467/24 + 555/25 + 687/145Cy5, Alexa 631, TOTO-3631QBS 405+493+575+653788/38QBP 425/30 + 514/30 + 592/25 + 709/100Confocal Laser Lines405nm, 488nm, 561nm, 640nmConfocal FiltersDetectorFiltersCH1SP470, SP545, SP620CH3LBF640, LP575, LP655, SP620Room Location March G.8 This article was published on 2026-05-26